NEUTRON · USER MANUAL · v1.4

Neutron CloudScaleLink and NS — user manual

Configuration, wiring, communication and weighing procedures for Neutron CloudScaleLink and Neutron NS.

Neutron AppEngine — custom weighing applications

Neutron AppEngine — JavaScript application development

JavaScript application development for CloudScaleLink Neutron and CloudScale Neutron NS: runtime, weighing, digital I/O, settings, storage, communication modules and application management. Manual 1.0 · Neutron AppEngine 1.0.0 · API 1 · Publication date: 16 November 2026.

1. Manual versions

  • January 2026, version 1: initial release.
  • March 2026, version 1.2: HTTPS support.
  • May 2026, version 1.3: connection procedures and wiring examples.
  • July 2026, version 1.4: multiple scale communication interfaces with firmware 1.7 or later; Modbus RTU and TCP with firmware 1.8 or later; country selection and adjustable Wi-Fi transmit power with firmware 1.8.3 or later.

2. Description

Neutron combines communication interfaces with optional local weighing and digital I/O. In CloudScaleLink mode, connect an existing scale, indicator or compatible serial device through UART, RS-232 or RS-485 and exchange data with Scale Monitor or other connected systems.

With the I/O board, Neutron provides four weighing channels through a multiplexed 24-bit ADC, up to 4,800 conversions per second, six digital inputs and six digital outputs. Wi-Fi and Bluetooth BLE are included; Ethernet requires the optional add-on.

CSL-NGP-WB — Wi-Fi and Bluetooth BLE, without Ethernet.

CSL-NGP-WBE — Ethernet, Wi-Fi and Bluetooth BLE.

3. Technical specifications

Check the installed CPU board, I/O board and Ethernet option before wiring or configuring the module. The specifications below distinguish communication electronics from the NS weighing assembly.

3.1. CPU board specifications

Supply the CPU board with 5–24 V DC ±5%. Typical consumption at 5 V is 80 mA with Wi-Fi only, rising to approximately 100 mA during scanning. With Ethernet, typical consumption is 170 mA, with peaks up to 250 mA.

  • UART / RS-232: selectable with SW1; RS-232 distance up to 15 m.
  • RS-485: address 0–255; distance up to 1,200 m with shielded 2 × 24 AWG twisted pair, outer braid and aluminium foil.
  • Serial parameters: 1,200, 2,400, 4,800, 9,600, 19,200, 38,400, 57,600 or 115,200 bit/s; 7 or 8 data bits; no, even or odd parity.
  • Bluetooth: BLE 5.0; bidirectional serial redirection to UART / RS-232 or RS-485.
  • Wi-Fi: 2.4 GHz, IEEE 802.11b/g/n; indicative range up to 50 m indoors or 150 m outdoors. Supported security: open, WPA2/WPA3-PSK and WPA2/WPA3-Enterprise with EAP-TLS or EAP-TTLS.
  • Ethernet: optional CS-NA-ETH; 10BaseT / 100BaseTX; automatic speed and full/half-duplex negotiation; link, speed, activity and duplex indicators.
  • Environment: −10 to +40 °C; maximum relative humidity 85%, non-condensing.
  • CloudScaleLink dimensions (L × W × H): CSL-NGP-WB 71 × 40 × 13 mm; CSL-NGP-WBE 74 × 40 × 22 mm.

3.2. I/O board specifications

The I/O board takes power directly from the CPU board and adds approximately 15 mA plus the connected I/O load. The four-channel, 24-bit sigma-delta ADC has programmable digital filtering and up to 4,800 conversions per second. Internal/displayable resolution is up to 22 bits, approximately 4.2 million internal divisions, depending on filtering and update rate. The channels are multiplexed: only one channel is converted at a time.

Configure up to four scales independently: a freely entered measurement unit, multi-division or multi-range operation with up to three ranges, division and maximum capacity for each range, zero tracking, automatic zero on startup, filtering, stability divisions and stability window. Calibration supports up to eight additional linearization points per scale.

  • Connect up to 16 analogue load cells of 350 Ω. Excitation: 5 V DC ±5%, maximum 250 mA.
  • Minimum signal per division: 0.023 µV. Minimum sensitivity: 1 mV/V; lower sensitivity can reduce the achievable resolution.
  • Four-wire connection: fit the reference jumpers shown in the connection diagram. Six-wire connection: use the sense/reference lines.
  • Scale interfaces: UART / RS-232, RS-485, BLE, cloud MQTT/WSS and HTTP/REST. Use the CSCP manual for commands and the protocol manuals for industrial integration.
  • Six optically isolated digital inputs: 10–24 V DC ±5%; input current 4–35 mA. ON/OFF state reporting or counting up to 1,000 Hz. Counter reports every 5 seconds; values can also be requested, set or reset by command. Up to four counters can operate simultaneously.
  • Maximum counter value: 18 446 744 073 709 551 615 (2⁶⁴ − 1).
  • Six PhotoMOS outputs: 5–24 V DC, maximum 150 mA. Control by command or weighing logic, including tolerance status and fast/slow dosing.
  • Neutron NS dimensions (L × W × H): CS-NS-WB 95 × 68 × 25 mm; CS-NS-WBE 95 × 68 × 35 mm.
  • Operating environment: −10 to +40 °C; maximum 85% relative humidity, non-condensing.

3.3. Dimensions

All dimensions are in millimetres. CloudScaleLink is the CPU-board assembly; Neutron NS combines the CPU and weighing I/O boards. Choose the drawing for the installed Ethernet option.

Table 1

Model

Dimensions (L × W × H), mm

CSL-NGP-WB

71 × 40 × 13

CSL-NGP-WBE

74 × 40 × 22

CS-NS-WB

95 × 68 × 25

CS-NS-WBE

95 × 68 × 35

CloudScaleLink: CSL-NGP-WB 71 × 40 × 13 mm; CSL-NGP-WBE with Ethernet 74 × 40 × 22 mm. Open the image to enlarge it.

Neutron NS: CS-NS-WB 95 × 68 × 25 mm; CS-NS-WBE with Ethernet 95 × 68 × 35 mm. Open the image to enlarge it.

3.4. CPU board controls and connectors

Locate the reset button, multi-purpose MD button, RGB status indicator, SW1 UART/RS-232 selector, screw terminal block for power and serial connections, optional adapter power socket and U.FL / IPEX connector for the external 2.4 GHz antenna.

CloudScaleLink processor board components and connectors. Open the image to enlarge it.

3.5. Connecting the module

Switch off Neutron and the connected device before wiring. Select the interface, set SW1 to match the connected device and wire the terminal block. After wiring, apply power and configure matching baud rate, data bits, parity, stop bits and protocol.

For RS-232, cross the transmit and receive lines and connect signal ground. For RS-485, connect A(+) and B(−), with signal ground where required by the installation.

3.5.1. Terminal block signals

  • RX: serial data input to Neutron.
  • TX: serial data output from Neutron.
  • B(−): negative differential RS-485 line.
  • A(+): positive differential RS-485 line.
  • GND: common signal ground.
  • Vdc+: positive DC supply input, 5–24 V DC.

Neutron NS terminal overview. RX and TX are the serial lines; B(−) and A(+) are the RS-485 lines; GND is ground and Vdc is power. Load-cell connections are on the right; digital inputs and outputs are on the left. When wiring, follow the individual terminal labels in the image and the relevant wiring diagram. Open the image to enlarge it.

3.5.2. UART, RS-232 and RS-485 wiring examples

Check the connected device's terminal assignments before wiring. RS-485 A/B naming can differ between manufacturers. If communication fails, first verify baud rate and protocol, then RS-232 TX/RX crossover or RS-485 polarity. The default serial settings are 9,600 bit/s, no parity and 8 data bits. Vdc+ is the external DC power input.

3.6. MD button

Each press of MD is acknowledged by the RGB indicator. Press the button the required number of times:

  • 1: restart the module.
  • 2: disable DHCP.
  • 3: enable the Wi-Fi access point.
  • 4: temporarily enable Bluetooth until the next reboot.
  • 5: restore factory settings.

The MD button can be disabled in General settings.

3.7. RGB LED indicator

  • Green: Wi-Fi or Ethernet connected.
  • Red: not connected; check Wi-Fi availability and signal or the Ethernet cable and module.
  • Yellow: Wi-Fi connection failed, for example due to credentials or encryption; with Ethernet, link established but waiting for a DHCP address.
  • Grey status: Wi-Fi disconnected or IP address lost.
  • Orange: unknown Wi-Fi status or Ethernet cable disconnected.
  • Flashing blue: Bluetooth debugging enabled, waiting for a connection.
  • Purple: MD button pressed.
  • Off: no power or module not operating.

4. Module ID and PIN

Each module has a unique module ID (MID) and a factory PIN printed on its label. Use them to identify and access the module. You can change the PIN in the configuration utility. A factory reset restores the factory PIN.

5. mDNS and hostname

Set the Device Name (DN) in General settings. Neutron converts it into a valid hostname; the mDNS name is that hostname followed by .local. If the device name is empty, MID is used.

  • Packing-Line-03 becomes packing-line-03.local.
  • Lab Scale A becomes lab-scale-a.local.
  • With an empty device name and MID T7BW2J, use t7bw2j.local.

Use the hostname for local access instead of entering the IP address.

6. Configuration utilities

  • BLE utility: browser-based Bluetooth configuration at apps.scale-monitor.com/bleNeutron.html.
  • Cloud Management Platform (CMP): remote configuration at login.scale-monitor.com; requires internet access over Wi-Fi or Ethernet.
  • Internal web server: local browser access through the module IP address or hostname. In access-point mode, use https://192.168.4.1 with HTTPS enabled. Match the scheme and port to the configured web-server settings.
  • CSLTools: Windows configuration over the local Wi-Fi/Ethernet network; available from apps.scale-monitor.com.

CMP supports all module settings. Availability in the other utilities depends on the particular function.

7. Communication interface

Bluetooth and Wi-Fi are enabled by default. In General settings, choose Wi-Fi or Ethernet for internet and LAN communication. Ethernet appears only when the optional Ethernet hardware is installed. Disable the network communication interface if the module will operate through Bluetooth only.

7.1. Multiple scale communication interfaces

With firmware 1.7 or later, enable several interfaces simultaneously under Scale communication interface, each with its own protocol and settings. Depending on hardware and firmware, the choices include Bluetooth, UART/RS-232, RS-485, TCP, WebSocket, HTTP/REST, Cloud Scales/MQTT, Cloud, Modbus TCP and Disabled.

  1. Open the configuration utility and Scale communication interface.
  2. Select the required interfaces.
  3. Configure the protocol and parameters for each interface.
  4. Save and restart the communication service or module when requested.

Examples: RS-232 plus Cloud Scales for a scale and PC/cloud connection; Modbus TCP over Ethernet/Wi-Fi for a PLC or SCADA; Modbus RTU over a serial line; Bluetooth for a local mobile application; TCP, WebSocket or HTTP/REST for custom software. Enable only the interfaces required for the installation.

8. Wi-Fi setup

Connect Neutron to a 2.4 GHz Wi-Fi network using the BLE tool or its access point. Prepare the SSID, credentials and, if needed, network settings and certificates.

  1. Select Wi-Fi as the connection interface.
  2. Enter the SSID or scan for and select the network.
  3. Select an open network, WPA2/WPA3 with a personal password, or Enterprise authentication.
  4. Enable DHCP to obtain network settings automatically. For a static configuration, enter the IP address, subnet mask, gateway and DNS.
  5. Save the settings and restart the device.

8.1. Wi-Fi setup through Bluetooth BLE

  1. On a Bluetooth-capable phone, tablet or PC, open the BLE configuration utility.
  2. Allow Bluetooth access, select the intended Neutron and choose Pair.
  3. Enter its PIN. A green connection indicator means connected; red means disconnected.
  4. Select Wi-Fi as the connection interface. Open Wi-Fi, enter or select the SSID and configure an open network, WPA2/WPA3 with a personal password, or Enterprise as described in this chapter. Select DHCP or enter the static IP address, subnet mask, gateway and DNS, then choose Save.
  5. Wait for confirmation and choose Reboot. The module connects using the saved settings.

On iPhone or iPad, use a browser with Web BLE support, such as Bluefy.

8.2. Wi-Fi setup through the internal web server

  1. Press MD 3 times to enable the access point.
  2. Connect your phone, tablet or PC to CS-XXXXXX, where XXXXXX is MID. The access-point password is the factory PIN; changing the module PIN does not change this password.
  3. Open https://192.168.4.1 or the module hostname with HTTPS enabled. If the web-server scheme or port has been changed, use its configured address.
  4. Log in with the current module PIN.
  5. Select Wi-Fi as the connection interface. Open Wi-Fi, enter or select the SSID and configure an open network, WPA2/WPA3 with a personal password, or Enterprise as described in this chapter. Select DHCP or enter the static IP address, subnet mask, gateway and DNS, then choose Save. Wait for confirmation that the settings were saved and restart the module.

If connection fails, reconnect through the access point and check the saved settings.

8.3. Scanning for Wi-Fi networks

Choose Scan and select a network from the results. Selecting a network also fills its BSSID; clear it if the installation requires roaming between access points with the same SSID. Scanning is unavailable while Neutron is trying to establish its Wi-Fi connection.

Signal strength is shown in dBm. A value closer to zero is stronger: −40 dBm is stronger than −70 dBm. Indicative levels: −30 to −50 excellent; −50 to −60 very good; −60 to −70 good; −70 to −80 weak; −80 to −90 very weak; below −90 unreliable.

8.4. Open and WPA2/WPA3-PSK networks

For an open network, enter SSID and set Encryption to None; leave the password empty. Open networks do not encrypt traffic and are suitable only where that limitation is acceptable, such as isolated testing.

For a password-protected network, enter SSID and Password and set Encryption to WPA2/WPA3. These networks use a pre-shared key and do not require a RADIUS server.

8.5. Static IP settings

With DHCP, the device obtains network settings automatically. For a static configuration, enter IP Address, Subnet Mask, Gateway and DNS.

8.6. WPA2/WPA3-Enterprise

Wi-Fi Enterprise and Ethernet 802.1X offer the same four authentication methods. Neutron acts as the client or supplicant. The access point or switch forwards authentication to the RADIUS server. RADIUS shared secret is configured between the access point or switch and RADIUS; do not enter it in Neutron.

  • EAP-TLS: Client certificate and matching private key; no user password is required.
  • EAP-TTLS: Username and password inside a TLS tunnel; PAP or MSCHAPv2.
  • PEAP-MSCHAPv2: Username and password inside a TLS tunnel; EAP-MSCHAPv2.
  • TTLS/PEAP auto: Neutron accepts TTLS or PEAP as requested by RADIUS. The inner-method selection applies to TTLS; PEAP uses EAP-MSCHAPv2.

Field descriptions

  • EAP identity: External identity presented by the device at the start of authentication.
  • Username: Username for inner authentication with TTLS and PEAP.
  • Password: Password for this user.
  • TTLS inner authentication: PAP or MSCHAPv2; in TTLS/PEAP auto this applies to TTLS.
  • RADIUS certificate DNS name: Optional name used to verify the server certificate.
  • Verify certificate dates: Enable or disable certificate validity-date verification.
  • Server CA certificate: CA used to verify trust in the RADIUS server certificate.
  • Client certificate: Neutron client certificate for EAP-TLS.
  • Client private key: Matching private key for EAP-TLS.

Certificates and server verification

  • CA certificate: confirms a trusted issuer.
  • Server certificate: used by the RADIUS server.
  • Client certificate: identifies Neutron with EAP-TLS.

Upload certificates and the key in PEM format. Use an unencrypted PEM private key that matches the client certificate. An empty upload field after opening the page does not mean the device has no certificate. Leaving the field untouched preserves the existing certificate.

Ethernet 802.1X requires a CA certificate. Wi-Fi can authenticate without a CA when name and date verification are both disabled. Wi-Fi requires a CA certificate to verify names or dates.

Optional name verification

Leaving RADIUS certificate DNS name empty disables name verification but preserves verification against the configured CA. If entered, the name is checked against the RADIUS server certificate. Find it in the server certificate under Subject Alternative Name → DNS Name, not in the CA or client certificate. Enter only the name, without https:// or a port, for example radius.example.com.

Date verification

When Verify certificate dates is enabled, the device clock must already be correct before authentication. NTP that becomes reachable only after login cannot set the clock for initial authentication through a closed network port. Disabling date verification preserves verification against the configured CA and server name. For compatibility, Wi-Fi defaults to date verification disabled. Existing saved settings are preserved on upgrade.

8.6.1. EAP-TTLS

  1. Select EAP-TTLS.
  2. Enter EAP identity, Username and Password.
  3. Under TTLS inner authentication, select PAP or MSCHAPv2 to match the RADIUS server configuration.
  4. Configure a CA certificate, or connect without a CA only with both name and date verification disabled.
  5. Optionally enter RADIUS certificate DNS name. Set Verify certificate dates, observing the CA requirements and ensuring that the device clock is correct.
  6. Select Save, wait for confirmation that the settings were saved, then restart the device with Reboot.

PEAP-MSCHAPv2

  1. Select PEAP-MSCHAPv2.
  2. Enter EAP identity, Username and Password.
  3. EAP-MSCHAPv2.
  4. Configure a CA certificate, or connect without a CA only with both name and date verification disabled.
  5. Optionally enter RADIUS certificate DNS name. Set Verify certificate dates, observing the CA requirements and ensuring that the device clock is correct.
  6. Select Save, wait for confirmation that the settings were saved, then restart the device with Reboot.

TTLS/PEAP auto

  1. Select TTLS/PEAP auto.
  2. Enter EAP identity, Username and Password.
  3. Select PAP or MSCHAPv2 for TTLS to match RADIUS. PEAP uses EAP-MSCHAPv2.
  4. Configure a CA certificate, or connect without a CA only with both name and date verification disabled.
  5. Optionally enter RADIUS certificate DNS name. Set Verify certificate dates, observing the CA requirements and ensuring that the device clock is correct.
  6. Select Save, wait for confirmation that the settings were saved, then restart the device with Reboot.

8.6.2. EAP-TLS

  1. Select EAP-TLS.
  2. Enter EAP identity.
  3. Upload the client certificate and its matching unencrypted PEM private key.
  4. Configure a CA certificate, or connect without a CA only with both name and date verification disabled.
  5. Optionally enter RADIUS certificate DNS name. Set Verify certificate dates, observing the CA requirements and ensuring that the device clock is correct.
  6. Select Save, wait for confirmation that the settings were saved, then restart the device with Reboot.

Troubleshooting

  • No connection: check the SSID, access point availability, signal strength and selection of the Wi-Fi connection interface.
  • No IP address: check DHCP or the static IP address, subnet mask, gateway and DNS.
  • Authentication fails: match the method and inner method to the RADIUS configuration.
  • Credential error: check the username and password.
  • Certificate error: check the correct CA, server name, client certificate and matching key.
  • Validity-date error: check the device clock and date-verification setting.
  • Changes are not active: confirm successful saving and restart the device.

TIMEOUT may also accompany a failed TLS exchange. Use logs from Neutron, the access point and RADIUS for diagnostics. Application status messages are visible at debug level Info or higher.

8.6.3. BSSID

BSSID is the MAC address of one specific access point. Enter it to restrict Neutron to that access point; scanning and selecting a network fills it automatically. Do not set BSSID when the module must roam between multiple access points sharing one SSID: a fixed BSSID prevents roaming and may cause weak or lost connectivity.

8.7. Country of use

Available with firmware 1.8.3 or later. In Wi-Fi settings, select the country where the device is physically installed, save and reboot if requested. The default is EU – Europe. Country selection controls permitted Wi-Fi channels and transmit-power limits. Do not select another region to obtain additional channels or power. Update this setting if the module is moved permanently to another country.

8.8. Wi-Fi transmit power

Available with firmware 1.8.3 or later. Use Auto unless the installation requires a specific value; firmware and country settings determine the permitted output. Selectable settings and approximate power: 20 dBm / 100 mW; 18 / 63; 15 / 32; 13 / 20; 11 / 13; 8.5 / 7; 7 / 5; 5 / 3; 2 / 1.6.

Higher power may improve range but increases consumption, interference and coverage outside the intended area. Reduce power when the access point is close, many modules share the area, or limited coverage and lower consumption are required. The selectable range is 2–20 dBm; actual permitted output also depends on country and channel. Configure Country of use first.

9. Ethernet

Ethernet requires the optional CS-NA-ETH module. The Ethernet menu is unavailable if the module is absent or not working.

  1. Connect the Ethernet-equipped device using a network cable.
  2. Select Ethernet as the connection interface.
  3. Enable DHCP to obtain network settings automatically. For a static configuration, enter the IP address, subnet mask, gateway and DNS.
  4. For a normal network without 802.1X, leave authentication disabled. For a protected network port, select the appropriate 802.1X method.
  5. Save the settings and restart the device.

9.1. 802.1X authentication

The methods and fields are the same as for Wi-Fi Enterprise in section 8.6. A CA certificate is mandatory for every Ethernet 802.1X method. See section 8.6 for certificates, name verification and date verification.

A) EAP-TLS

  1. Select EAP-TLS.
  2. Enter EAP identity.
  3. Upload the client certificate and its matching unencrypted PEM private key.
  4. Upload the CA certificate.
  5. Optionally enter RADIUS certificate DNS name. Set Verify certificate dates, observing the CA requirements and ensuring that the device clock is correct.
  6. Select Save, wait for confirmation that the settings were saved, then restart the device with Reboot.

B) EAP-TTLS

  1. Select EAP-TTLS.
  2. Enter EAP identity, Username and Password.
  3. Under TTLS inner authentication, select PAP or MSCHAPv2 to match the RADIUS server configuration.
  4. Upload the CA certificate.
  5. Optionally enter RADIUS certificate DNS name. Set Verify certificate dates, observing the CA requirements and ensuring that the device clock is correct.
  6. Select Save, wait for confirmation that the settings were saved, then restart the device with Reboot.

C) PEAP-MSCHAPv2

  1. Select PEAP-MSCHAPv2.
  2. Enter EAP identity, Username and Password.
  3. EAP-MSCHAPv2.
  4. Upload the CA certificate.
  5. Optionally enter RADIUS certificate DNS name. Set Verify certificate dates, observing the CA requirements and ensuring that the device clock is correct.
  6. Select Save, wait for confirmation that the settings were saved, then restart the device with Reboot.

D) TTLS/PEAP auto

  1. Select TTLS/PEAP auto.
  2. Enter EAP identity, Username and Password.
  3. Select PAP or MSCHAPv2 for TTLS to match RADIUS. PEAP uses EAP-MSCHAPv2.
  4. Upload the CA certificate.
  5. Optionally enter RADIUS certificate DNS name. Set Verify certificate dates, observing the CA requirements and ensuring that the device clock is correct.
  6. Select Save, wait for confirmation that the settings were saved, then restart the device with Reboot.

Ethernet authentication states

  • DISABLED: Authentication is disabled.
  • WAIT_LINK: Waiting for the physical link.
  • AUTHENTICATING: Authentication is in progress.
  • AUTHORIZED: Authentication succeeded.
  • REJECTED: Authentication failed.
  • TIMEOUT: Authentication did not complete within the expected time.
  • CONFIG_ERROR: Invalid or incomplete configuration, or an initialization error.
  • IO_ERROR: Error transferring authentication frames.

A physical link and an assigned IP address alone do not prove successful 802.1X authentication. AUTHORIZED confirms successful authentication.

Troubleshooting

  • No connection: check the network cable, switch port and selection of the Ethernet connection interface.
  • No IP address: check DHCP or the static IP address, subnet mask, gateway and DNS.
  • Authentication fails: match the method and inner method to the RADIUS configuration.
  • Credential error: check the username and password.
  • Certificate error: check the correct CA, server name, client certificate and matching key.
  • Validity-date error: check the device clock and date-verification setting.
  • Changes are not active: confirm successful saving and restart the device.

TIMEOUT may also accompany a failed TLS exchange. Use logs from Neutron, the switch and RADIUS for diagnostics. Application status messages are visible at debug level Info or higher.

10. Serial connection

Serial Port 1 operates as UART or RS-232; Serial Port 2 provides RS-485. Use either port to bridge an existing serial device, or to communicate with the local weighing electronics.

For each port, set Serial redirect to Disabled, Cloud, WebSocket, Bluetooth, HTTP or TCP. To communicate with the scale on the I/O board through a serial port, leave its redirection disabled.

10.1. UART / RS-232

Serial Port 1 defaults to RS-232. Configure redirection, baud rate, data bits and parity. Available baud rates are 1,200, 2,400, 4,800, 9,600, 19,200, 38,400, 57,600 and 115,200 bit/s; choose 7 or 8 data bits and no, even or odd parity. Select CR, LF, CRLF or up to two custom ASCII characters as the string terminator. Choose whether to append that terminator automatically when forwarding data to the serial port.

10.1.1. Switching between RS-232 and UART

Switch off Neutron before changing SW1. Match SW1 to the electrical interface of the connected device. Connecting a UART device while SW1 is set to RS-232 can damage that device; connecting an RS-232 device while SW1 is set to UART can damage Neutron.

10.1.2. UART / RS-232 terminal wiring

  • Connect Neutron RX to the external device TX.
  • Connect Neutron TX to the external device RX.
  • Connect Neutron GND to the external device signal ground.

Confirm the device pinout and SW1 position before applying power.

10.2. RS-485

Use Serial Port 2 for differential RS-485 communication, including a shared bus with several devices and one master. Supported distance is up to 1,200 m with the specified cable and suitable installation conditions. Neutron controls switching between transmission and reception automatically.

Configure redirection, RS-485 address (0–255), baud rate, 7/8 data bits, parity and terminator. Baud rates and terminator choices are the same as UART/RS-232; automatic terminator appending is optional.

10.2.1. RS-485 wiring

Connect Neutron A(+) to the external device or bus A(+), and B(−) to B(−). Connect signal ground where required. Check the manufacturer's polarity convention if A/B names differ.

10.3. Modbus RTU

Available with firmware 1.8 or later. Under Scale communication interface, enable RS-232 or RS-485 for Modbus RTU. Match baud rate, data bits, parity, stop bits and Modbus slave address / Unit ID to the master. Set the address with MBA; the default is 1.

For RS-232, connect TX to RX, RX to TX and signal ground; check DTE/DCE pin assignments. For RS-485, connect A(+), B(−) and ground where needed. Use a daisy-chain bus, not star wiring, and terminate only at the two physical ends where termination is required.

If communication fails, check A/B polarity, serial parameters and Unit ID. Open the Modbus RTU/TCP manual for the register map, functions, commands, data formats and calibration interface.

11. Cloud connection through MQTT

In Cloud settings, enable the connection and select its transport. If Cloud is disabled, Neutron does not connect to the broker and remote cloud management is unavailable.

  • MQTT: unencrypted TCP; use only where an unencrypted connection is acceptable.
  • WSS: WebSocket with TLS encryption.
  • MQTTS — Username & Password: TLS with broker credentials; choose Custom credentials for your own broker.
  • MQTTS — Client Certificate (mTLS): TLS with a client certificate and private key; set Credentials to None.

For WSS or MQTTS, provide the appropriate CA certificate when the broker uses a self-signed certificate or an authority not already trusted. A valid trusted chain is needed to verify broker identity.

Enter Server (hostname or IP), Port and Client ID; the default Client ID is MID. Credentials options are Built-in for Scale Monitor, Custom for your username/password, and None when no username/password is required or mTLS is used.

  • QoS 0: at most once, no acknowledgement; default and lowest overhead.
  • QoS 1: at least once, acknowledged; duplicates are possible.
  • QoS 2: exactly once at the MQTT delivery level; higher processing and bandwidth cost.

For unreliable links, consider QoS 1. Use QoS 2 only when required rather than for continuous high-rate communication. Group topic and Group topic without SN/MID are manufacturer settings.

11.1. MQTT topics

Replace MID with the module ID. Neutron subscribes to RX topics for commands and publishes data or responses to TX topics; it does not publish commands on RX or subscribe to TX.

  • $CSL/MID/RX / $CSL/MID/TX: UART / RS-232.
  • $CSL/MID/RX/S1 / $CSL/MID/TX/S1: RS-485.
  • $CSL/MID/RX/SC / $CSL/MID/TX/SC: local scale.
  • $CSL/MID/RX/IO / $CSL/MID/TX/IO: digital I/O commands and responses.
  • $CSL/MID/STATUS: connection status. Connected message: STA;S=C;MID=XXXXXX. Broker Last Will on disconnection: STA;S=D;MID=XXXXXX.
  • $CSL/MID/DEBUG: communication debug messages.

For Wi-Fi startup and connection diagnostics, use Bluetooth debugging.

11.2. MQTT debugging tool

Open apps.scale-monitor.com and select MQTT client. Connect to the intended broker, subscribe to the module's TX topics and publish test commands to the corresponding RX topics.

11.3. AWS IoT Core

Connect Neutron to AWS IoT Core with MQTT over TLS and X.509 client-certificate authentication. Prepare a Thing, its certificate and private key, a suitable IoT policy and the account's device endpoint.

11.3.1. Creating a Thing

  1. Open AWS IoT Core and start the device-connection procedure.
  2. Copy the device endpoint for the Neutron Server setting.
  3. Create a Thing; using MID as Thing name simplifies matching Client ID.
  4. In the connection wizard, choose the offered device platform and Node.js SDK option where requested.
  5. Download the connection kit containing the device certificate and private key. Keep the private key confidential.

11.3.2. Configuring the IoT policy

In IoT Core, open Security → Policies and edit the policy attached to the device certificate. Allow iot:Connect for the intended Client ID, iot:Publish and iot:Receive for the required topics, and iot:Subscribe for their topic filters. Restrict resources to your account, region and module.

In the AWS connection mode, the device example uses CSL/MID/… topics. Include the required AWS topics only for the functions you use. Set the edited policy version active and save it. Do not copy another account's ARN or use unrestricted * resources for a production installation.

In this template, replace REGION, ACCOUNT_ID and MID with your region, account and device identifier. Add only the additional AWS topics required by your application.

{
  "Version": "2012-10-17",
  "Statement": [
    {
      "Effect": "Allow",
      "Action": "iot:Connect",
      "Resource": "arn:aws:iot:REGION:ACCOUNT_ID:client/MID"
    },
    {
      "Effect": "Allow",
      "Action": [
        "iot:Publish",
        "iot:Receive"
      ],
      "Resource": "arn:aws:iot:REGION:ACCOUNT_ID:topic/CSL/MID/*"
    },
    {
      "Effect": "Allow",
      "Action": "iot:Subscribe",
      "Resource": "arn:aws:iot:REGION:ACCOUNT_ID:topicfilter/CSL/MID/*"
    }
  ]
}

11.3.3. Neutron settings for AWS

  1. Connect through the BLE utility or the internal web server and open Cloud.
  2. Select MQTTS — X.509 Client certificate (mTLS).
  3. Upload the certificate ending in cert.pem and its corresponding key ending in private.key.
  4. Enable Connecting to Amazon Web Services cloud (AWS).
  5. Enter the copied endpoint as Server and port 8883.
  6. Match Client ID to the Thing name; set Credentials to None.
  7. Save and reboot.

If connection fails, use Bluetooth debugging to inspect the connection messages.

11.3.4. Testing AWS communication

In IoT Core, open Test → MQTT test client. Replace MID with the module ID, subscribe to CSL/MID/TX and publish a command to CSL/MID/RX. Check the response on TX. For other interfaces, use their corresponding topic suffixes.

12. Microsoft Azure IoT Hub

Neutron supports Azure IoT Hub over MQTTS with SAS-token authentication or the supported X.509 mode. In this implementation, use only one cloud bridge at a time. When several are enabled, the processing priority is scale, then RS-232, then RS-485; lower-priority bridges are ignored.

  1. In Azure, create or choose an IoT Hub and open Devices.
  2. Add a device. Use MID as the device identifier/Client ID, or remember the custom identifier for the Neutron configuration. For SAS authentication, choose Symmetric key and automatic key generation; enable connection to the hub.
  3. Open the created device and retrieve its hub hostname and primary key.
  4. In Neutron Cloud settings, select MQTTS — Username & Password, enter the hub hostname, port 8883 and matching Client ID.
  5. Select Azure Cloud — SAS token under Credentials and enter the primary key.
  6. Save and reboot. Use debugging to check the connection status.

If using the X.509 option instead, select the matching MQTTS X.509 mode and upload the required certificates and key.

12.1. Testing Azure communication

Use Azure IoT Explorer. Open the device, start telemetry monitoring and send a cloud-to-device message containing the command for Neutron. Check the responses in the telemetry view.

13. WebSocket

Use WebSocket for bidirectional communication with a remote server. For serial bridging, choose WebSocket under Serial redirect. For the local scale, select WebSocket under Scale communication interface. Both WS and encrypted WSS are supported, with optional HTTP Basic Authentication.

13.1. WebSocket quick configuration

  • Local unencrypted test: WS, server IP/hostname, port 80 or the configured port, path such as /ws; no TLS certificate.
  • Local self-signed TLS test: WSS, TLS port such as 443, configured path; use the test-only insecure option only when certificate validation is deliberately disabled.
  • Production/cloud: WSS, server domain, port 443 or the configured TLS port, configured path and trusted CA certificate; keep insecure mode disabled.
  • Authenticated backend: use the production settings and enable Basic Authentication with the required username and password.

13.2. WebSocket mode

Disabled prevents a WebSocket connection. WS uses unencrypted TCP and is intended for trusted local networks. WSS uses TLS encryption and is the appropriate choice for public networks and cloud servers.

13.3. Insecure TLS option

In WSS mode, set Allow insecure SSL connection to No to validate the server certificate against the configured CA. Yes accepts certificates without validation and significantly reduces security; use it only for controlled testing or troubleshooting.

13.4. CA certificate

Upload the CA certificate in PEM format to verify the server's TLS certificate. It is required for WSS when insecure connections are not allowed.

13.5. Server

Enter the WebSocket server hostname or IP address, for example ws.example.com or 192.168.1.100.

13.6. Port

Enter the server's TCP port: typically 80 for WS, 443 for WSS, or the custom port configured on the server.

13.7. Path

Enter the WebSocket endpoint path, such as /, /ws or the path required by the server. An empty field uses the root path /.

13.8. Basic Authentication

Enable Use Basic Authentication when the server requires a username and password during the WebSocket handshake. Enter both credentials. If disabled, these credentials are not sent.

14. Web server and REST API

The internal web server provides module configuration and REST bridging to serial ports or the local scale. HTTPS is enabled by default on port 443 with a self-signed certificate. Log in with the module PIN. HTTPS can be disabled in HTTP settings; setting the port to 0 disables the web server completely.

Use the configured scheme, IP/hostname and port, for example https://192.168.50.79, https://t7bw2j.local, or http://192.168.50.79:8080 when HTTP is explicitly configured on port 8080. A self-signed certificate may trigger a browser warning until its CA is trusted and the hostname matches.

14.1. TLS certificate

  1. Open HTTP settings.
  2. Upload your TLS certificate and its matching private key in the supported format.
  3. Reboot to activate the certificate for HTTPS connections.

Use an organizational or trusted CA certificate where required by your installation.

14.1.1. Trusting the self-signed certificate

  1. On the login page, choose Install CA certificate and install the CA on the client operating system or browser.
  2. Use a hostname matching the certificate's *.neutron.local pattern, for example abc123.neutron.local.
  3. Ensure that hostname resolves to the module IP through your DNS configuration or an explicit hosts-file entry.

Both certificate trust and hostname matching are required to avoid validation warnings.

14.2. REST bridge

For serial access, set Serial redirect to HTTP. For the I/O-board scale, set its communication interface to HTTP. Otherwise the corresponding endpoint returns 404.

  • S1: UART / RS-232.
  • S2: RS-485.
  • SC: local scale.

The module waits up to 5 seconds for a device response and returns it in the HTTP body. If nothing arrives, the body contains [no serial response] or [no scale response].

14.2.1. REST token

Set an optional token in the HTTP tab. When enabled, include the tok parameter in GET and POST requests. The maximum token length is 128 ASCII characters. Percent-encode reserved characters when constructing a URL, especially &, = and #; a raw # starts a fragment and is not sent to the server.

Example with placeholders: https://MODULE-IP/SC/?cmd=SXI&tok=YOUR-ENCODED-TOKEN. Keep your actual token confidential.

14.2.2. GET requests

Pass the command in cmd, for example https://MODULE-IP/S1/?cmd=data. Use S2 or SC for the other interface and add tok when token protection is enabled.

14.2.3. POST requests

Send the data in the POST body, either as plain data or through the cmd parameter. The module forwards it to the selected serial port or scale endpoint. Include tok if configured.

15. Bluetooth

Enable or disable Bluetooth in its settings. To bridge a serial port, select Bluetooth under that port's Serial redirect setting. Configure the string terminator where needed.

15.1. Bluetooth BLE GATT

Use these UUIDs in a BLE application:

  • Service: 49535343-fe7d-4ae5-8fa9-9fafd205e455.
  • Write characteristic: 49535343-8841-43f4-a8d4-ecbe34729bb3.
  • Notify characteristic: 49535343-1e4d-4bd9-ba61-23c647249616.
  1. Scan for devices advertising the service.
  2. Connect to Neutron and open the primary service.
  3. Write commands to the Write characteristic.
  4. Subscribe to the Notify characteristic for responses.

15.2. Bluetooth debugging

In General settings, set Debug interface to Bluetooth and reboot. A flashing blue RGB LED means Neutron is waiting for a Bluetooth connection before starting Wi-Fi, Ethernet, I/O and other interfaces, so startup messages can be captured.

The wait is up to 30 seconds. If no client connects, normal startup continues; debugging messages remain available after a later connection. Use the BLE utility and its Bluetooth tab. A local BLE connection also supports diagnostics when Neutron has no internet connection. Remote support additionally depends on connectivity of the device running the utility.

15.3. Testing serial redirection

Set the selected serial port's redirection to Bluetooth. In the BLE utility's Bluetooth tab, enter data in Command and choose Send. Neutron forwards it to the serial device; replies appear in Received.

16. TCP

In the TCP tab, select Server, Client or Server & Client. Configure each enabled mode separately.

16.1. Server mode

Neutron listens for incoming connections on its configured TCP port. The default is 10010. The external system connects to the module's IP address and this port.

16.1.1. Exclusive TCP server connection

With TCP server port exclusive enabled, an existing connection causes new connection attempts to be rejected. With it disabled, a new connection replaces and disconnects the existing one.

16.2. Client mode

Enter the destination IP address and TCP port. Neutron initiates the connection and automatically attempts to reconnect after connection loss, for example following a Wi-Fi interruption or server outage.

16.3. TCP bridge to cloud

Enable Bridge to cloud for the TCP server, client or both. Data received through TCP is forwarded to the cloud, and cloud data is sent back through TCP. Configure server and client redirection independently. Use this for TCP-capable equipment such as network scales and label printers. Application-level bridge details are available with the CSL API documentation.

16.3.1. Connecting a network label printer

  1. Set TCP mode to Client.
  2. Enter the printer IP and TCP port; a common printer port is 9100. Obtain the actual address and port from the printer configuration.
  3. Enable the client bridge to cloud.
  4. In Scale Monitor Settings → Label Printer, enable Use bridge mode.
  5. If one CloudScaleLink serves both the scale and printer, use the same MID and PIN configured for that scale connection.

16.3.2. Zebra printer confirmations and alerts

  1. Set Neutron TCP mode to Server & Client and enable Bridge to cloud for the server.
  2. Open the Zebra printer's web interface and Alert Setup.
  3. Choose Add Alert Message and destination TCP.
  4. Enter Neutron's IP address and its TCP server port, default 10010.
  5. Add the required messages and authenticate with the printer's current credentials. Some models use the factory default 1234; confirm the credentials for your printer.
  6. Choose Save Printer Settings so the configuration survives a printer reboot.

The printer can then send supported print confirmations and alerts, such as out of labels, no ribbon or open cover, back to Scale Monitor.

16.4. Modbus TCP

Available with firmware 1.8 or later. Modbus TCP provides PLC, SCADA, HMI, MES and custom software with access to weighing values, scale state, I/O, commands and supported calibration data.

  1. Configure and verify Wi-Fi or Ethernet.
  2. Enable Modbus TCP under Scale communication interface.
  3. Set Unit ID with MBA, default 1.
  4. Save and restart the communication service or module when requested.
  5. Configure the master for the module IP, TCP port 502 and the selected Unit ID.

Use Ethernet where a wired connection is available; check Wi-Fi latency under actual operating conditions. Modbus TCP can run alongside other enabled interfaces, such as a serial scale connection. Open the Modbus RTU/TCP manual for the shared register map and command handling. For OPC UA integration, use the OPC UA manual.

17. I/O board

The I/O board adds four multiplexed 24-bit weighing channels, six digital inputs and six digital outputs. Inputs can count up to 1,000 Hz. Configure digital I/O using the BLE utility or CMP. Use the appropriate communication API for commands, input states and counter values.

17.1. Digital inputs

Open the I/O dashboard to view current input and output states. Choose an input's gear icon and select ON/OFF or Counter mode.

17.1.1. ON/OFF mode

Set the debounce filter: the signal must remain stable for this time before a change is accepted. Typical settings are 5–100 ms.

17.1.2. Counter mode

Use inputs 1–4 for counters. Set the glitch filter to reject noise pulses shorter than the selected duration; typical settings are 1–20 µs. This filter applies only in Counter mode. Up to four counters can be active simultaneously.

17.2. Digital outputs

Change outputs manually through commands, BLE or CMP, or assign an automatic function. Open the output gear icon, select the required function and save. Automatic functions can follow weighing tolerance or dosing states.

18. Scale

The I/O board provides four independently configurable weighing channels through a multiplexed 24-bit ADC, with up to 4,800 conversions per second. Conversion takes place on one channel at a time.

Depending on configuration, communicate through UART/RS-232, RS-485, BLE 5.0, MQTT over TCP or secure WSS, WebSocket WS/WSS, or HTTP/REST.

Open the CSCP manual for scale commands, Modbus RTU/TCP for registers and PLC communication, or OPC UA for the server model and integration.

18.1. Calibration display

  • Status: = stable, ~ unstable, UL underload, OL overload.
  • Range: active weighing range, 1–3.
  • Channel: selected ADC channel, 1–4.
  • ADC Value: raw converter value.
  • Millivolts: measured load-cell signal in mV.

The calibration menu provides Tare, Clear tare, Zero and Temperature. TEMP displays the ADC internal temperature. Weighing conversion is temporarily suspended while temperature is measured.

18.2. Channel switching

Select the Channel field and choose the required channel. Its settings are loaded automatically.

18.3. Metrological settings

For the selected channel, enter the measurement Unit, such as kg, g, t, mg or lb, or your own unit text. Select one to three weighing ranges and the required Multi-division or Multi-range mode. Set the divisions d1–d3 and maximum capacities Max1–Max3 for the corresponding ranges.

18.4. Scale parameters

  • Scale interface: communication interface; for Scale Monitor, select Cloud, RS-232 or Bluetooth to match the intended connection.
  • Zero tracking: Disabled or ¼, ½, 1, 2, 4, 8 or 10 divisions.
  • Boot on zero: percentage of maximum capacity that can be zeroed automatically at startup.
  • Zero range: permitted zero-command range as a percentage of maximum capacity.
  • Filter: digital filtering; 0 disables filtering and permits the maximum conversion rate of 4,800 Hz. Higher values improve stability but increase settling time.
  • HR filter: additional high-resolution noise filtering with a longer settling time.
  • Number of divisions for stability: permitted weight variation for a stable state. With 0, weight is always treated as stable.
  • Command timeout: maximum wait for a command to execute. For example, a 2-second tare timeout cancels the command if stability is not reached within 2 seconds.

18.5. Calibration

Calibrate zero first. Then add up to eight linearization points for the selected scale.

18.5.1. Automatic calibration

For each point, enter the mass of the load to be placed on the receptor, apply that load and choose Calibrate. Increase the load for every subsequent point. A lower load than the previous point decreases the ADC value and causes a calibration error.

18.5.2. Calibration status messages

Read the Net field: CS means calibration starting; C1–C10 counts attempts while waiting for a stable signal; CF means completed successfully; CE means an error.

If stability is not reached after 10 attempts, repeat the point under more stable conditions. The option Do not check stability bypasses stability verification and can store an unstable point; use it only when that behaviour is intentional.

18.5.3. Manual calibration

For manual calibration or fine adjustment, enter the ADC values and calibration factors directly.

18.6. Advanced settings

The stability window adds a time check after the divisions-based stability condition has been reached:

  • Quick: 0.25 seconds.
  • Medium: 0.5 seconds; default.
  • Slow: 1 second.

If weight changes during the window, stable status is not released. Use Medium or Slow for high-resolution scales; Quick can suit standard-resolution scales up to 10,000 divisions in stable conditions. Wait ADC is reserved for manufacturer use and should remain disabled during normal operation.

19. Cloud Management Platform (CMP)

Register at register.scale-monitor.com and sign in to CMP. Manage activated modules with cloud management enabled and an internet connection. CMP provides connection status, Wi-Fi signal, connectivity tests, LED identification, remote reboot, configuration and communication debugging.

19.1. Module list

The list shows modules activated in your CMP account. Choose Ping all modules to request their current status; offline modules are indicated. Open an individual module for connection tests, configuration or debugging.

19.2. Module activation

  1. In CMP, open Menu → CloudScaleLink → Modules → Activate module.
  2. Enter the module MID and PIN from its label, or the current PIN if changed.
  3. Choose Activate module.

Adding a scale in Scale Monitor also activates its module in CMP. Once activated in your account, another account cannot activate the same module until you deactivate it.

19.3. Module deactivation

Choose Deactivate module, enter the requested module details and confirm. The module is removed from your account's list and can then be activated in another account.

19.4. Module settings

Choose Settings. If the module is online, CMP establishes the configuration connection. Change the required settings, such as IP configuration or DHCP, and save. Reboot when the selected setting requires it.

19.5. Toolbar controls and status

  • Cloud: green means connected, red disconnected.
  • Wi-Fi: green indicates excellent signal, orange good and red poor; a crossed red icon indicates no cloud connection.
  • Ethernet: green means connected, red means no Ethernet connection.
  • Ping: request a response to check that the module is online.
  • Identification/light: toggle the module's green LED to locate the device; choose again to stop.
  • Reload data: refresh module data from the cloud.
  • Debug: open the communication debugger.

19.6. Debugging through CMP

The debugger displays communication between the module, serial device and connected software/cloud. Sent commands appear green, received messages orange and errors pink. Enter a command in the command panel and send it to inspect the device response.

19.6.1. Managing debug commands

Choose Add to create a command field, enter the command and choose its Send button. Use X beside the field to delete that command.

20. System resources

Open System resources to monitor uptime, CPU load, memory and the counts of Wi-Fi/Ethernet disconnections and lost IP addresses since startup.

21. Remote assistance

Connect to Neutron through the BLE utility on your phone or another Bluetooth-capable device and choose Request virtual assistance. When support accepts the request, use the chat while the assistant inspects and diagnoses the module. The device running the utility needs the connection required to reach support.

Configuration access is blocked during assistance to avoid conflicting changes and is restored when the session ends.

22. Troubleshooting

Check power, wiring, network status and the relevant interface settings before changing several settings at once. Use Bluetooth debugging for startup and network diagnostics.

22.1. No Wi-Fi connection

  • Check the exact SSID and password. Leave the password empty for an open network.
  • Check BSSID if configured; clear it where roaming is required.
  • Check that the network is available and signal strength is sufficient.
  • Use a 2.4 GHz network; 5 GHz is not supported.
  • Check the external antenna and its connector. Wi-Fi operation requires the external antenna.

22.2. Wi-Fi scanning is unavailable

Scanning cannot run at the same time as a Wi-Fi connection attempt. If Wi-Fi is selected as the communication interface and the module is still trying to connect, scanning is temporarily unavailable.

22.3. No CMP connection

  1. Check the Wi-Fi or Ethernet connection. Green RGB status means connected; red means disconnected. With Ethernet and DHCP, yellow means waiting for an IP address.
  2. Verify a valid IP address, subnet and gateway.
  3. Check DNS. If DNS is unavailable, use the intended server IP instead of its hostname where supported.
  4. Check internet access and the cloud-status indicator.
  5. Ensure the network permits the required outbound connections and does not block the configured service ports.
  6. Use Bluetooth debugging to inspect the actual connection error.

Frequently asked questions

Which password does the Wi-Fi access point use?

Use the factory PIN printed on the module label. Changing the module PIN does not change the access-point password.

Are all four weighing channels converted simultaneously?

No. The ADC is multiplexed and converts one channel at a time. Configure the channels separately.

Where are the communication commands and registers?

Use the CSCP manual for scale commands, the Modbus RTU/TCP manual for its register map, and the OPC UA manual for the server model. Direct links are in chapter 18 and the protocol cards.