Inspection body and legal verification
AD LAB holds accreditation K-007 to SIST EN ISO/IEC 17020. Inspection and legal verification are related but distinct: verification also requires MIRS designation for the relevant scope.
AD LAB · K-007
Arrange inspection, legal verification and documentation. Coordinate preparation, implementation and the next due date for mass instruments and blood pressure monitors.
AD LAB holds accreditation K-007 to SIST EN ISO/IEC 17020. Inspection and legal verification are related but distinct: verification also requires MIRS designation for the relevant scope.
For verification, the instrument must be technically sound, clean and ready. Check documentation, markings and condition; necessary service is separate from inspection or verification.
After successful verification, conformity is confirmed by the prescribed mark or POS. If you also require a certificate of conformity with regulations, specify it when ordering. Link the documents to a Soter equipment record.
AD LAB
Accredited activities are performed by AD LAB, metrološki laboratorij d. o. o. Scope, locations and restrictions are specified in annexes K-007 and LK-014, effective from 14 May 2026.
| Instrument or service | Scope and conditions |
|---|---|
| Non-automatic weighing instruments | Periodic and extraordinary inspection: class I up to 5 kg; II up to 120 kg, or 200 kg where n < 12,000; III and IIII up to 100 t. Where n > 3,000, inspection is at the place of use only, except instruments with automatic adjustment. |
| Automatic gravimetric filling and catchweighing instruments | Periodic and extraordinary inspection at the place of use. |
| Weighing instruments for construction | Periodic and extraordinary inspection at the place of use within the scope of the Tupaliče and Maribor locations. |
| Road vehicles in motion | Initial, periodic and extraordinary inspection at the place of use. |
| Weights | Tupaliče: initial, periodic and extraordinary inspection; F2 and M1 from 100 mg to 20 kg, M2 from 5 g to 20 kg. Skoke: periodic and extraordinary inspection of F2, M1 and M2 from 100 mg to 20 kg. Only in the inspection body laboratory. |
| Blood pressure measuring instruments | Mechanical, mercury and electronic instruments: periodic and extraordinary inspection at permanent premises and on site. |
Full K-007 scope annex (PDF) · Slovenian Accreditation register
This is the scope of accredited inspections. Statutory verification must also fall within the valid MIRS designation.
PLAN AHEAD
| Instrument or service | Scope and conditions |
|---|---|
| Non-automatic weighing instruments up to and including 9,000 kg | 2 years |
| Non-automatic weighing instruments above 9,000 kg | 1 year |
| Automatic catchweighing instruments | 1 year |
| Automatic gravimetric filling instruments | 1 year |
| Discontinuous totalising automatic weighing instruments | 2 years |
| Continuous totalising automatic weighing instruments | 2 years |
| Automatic rail weighbridges | 2 years |
| Weighing instruments for construction | 2 years |
| Automatic instruments for weighing road vehicles in motion | 1 year |
| Weights from 1 mg to 50 kg | 1 year; class M2: 2 years |
| Weights from 50 kg to 5,000 kg | 1 year |
| Blood pressure measuring instruments | 2 years |
This overview of general intervals does not mean every listed instrument type is within AD LAB's scope. See K-007 for the service scope.
Legal basis:Slovenian rules on periodic verification intervals for measuring instruments and regulations for individual instrument types.MIRS verification intervals (PDF).
FREQUENTLY ASKED QUESTIONS
A prescribed examination and test checking whether an instrument meets metrological requirements. Successful verification is confirmed by a verification mark or certificate of conformity where the conditions for issue are met.
MIRS · regulated measuring instruments · Slovenian rules on verification procedures for measuring instruments · Metrology Act (ZMer-1)
Legal metrology regulates measurements important to public interest and fair trade. In Slovenia, verification is performed by the Metrology Institute (MIRS) and bodies it designates. MIRS inspectors conduct metrological supervision. Accreditation alone is not designation to perform verification.
MIRS · regulated measuring instruments · Slovenian rules on verification procedures for measuring instruments · Metrology Act (ZMer-1)
Intended use is decisive: examples include trade by weight, calculation of levies, official proceedings, patient weighing, pharmacy preparation of medicines and sale of prepacked goods. Slovenian requirements derive from the Metrology Act (ZMer-1) and the relevant instrument regulations. Tell us how the result is used.
MIRS · regulated measuring instruments · Slovenian rules on verification procedures for measuring instruments · Metrology Act (ZMer-1)
A regulated scale is used for a purpose covered by metrological legislation and must meet conformity-assessment and verification requirements. An internal process scale serves a process outside that regulated use. Its suitability is managed through the process and quality system, usually with calibration and interim checks. An industrial location alone does not determine the category.
MIRS · regulated measuring instruments · Slovenian rules on verification procedures for measuring instruments · Metrology Act (ZMer-1)
Check the nameplate, type, serial number, prescribed markings and conformity documentation from before first use. For newer non-automatic instruments intended for regulated use, check CE and supplementary metrology markings; older instruments follow their applicable procedure. CE marking alone is insufficient. Send a nameplate photo and available documents. Periodic verification cannot simply convert an instrument without the required conformity into one suitable for regulated use.
MIRS · regulated measuring instruments · Slovenian rules on verification procedures for measuring instruments · Metrology Act (ZMer-1)
A certificate of conformity of an instrument with regulations (POS) documents conformity during legal verification. It is issued where the instrument's construction or use makes a verification mark impractical. You may also request one for a marked instrument, for records or audit evidence; an additional charge may apply. POS differs from the manufacturer's EU declaration of conformity and does not replace a calibration certificate. The basis is Article 10 of the Slovenian Rules on the procedure for verification of measuring instruments.
MIRS · regulated measuring instruments · Slovenian rules on verification procedures for measuring instruments · Metrology Act (ZMer-1)
It depends on instrument type and, for non-automatic scales, maximum capacity. See the interval table below. Where the interval is under five years, validity runs to the end of the month in which it expires. Intervention, damaged seals or other prescribed circumstances may require earlier extraordinary verification.
MIRS · regulated measuring instruments · Slovenian rules on verification procedures for measuring instruments · Metrology Act (ZMer-1)
Yes. Alongside mass instruments, we arrange verification of blood pressure monitors through AD LAB's services. Provide manufacturer, model, serial number and quantity; suitability is checked against the applicable instrument regulations.
MIRS · regulated measuring instruments · Slovenian rules on verification procedures for measuring instruments · Metrology Act (ZMer-1)